1. Software Issues

1.4 System Message: “System Recovering. Please wait”

Click “CGM” button on Home Screen and check current value
    If Current is above 50 nA: Check if transmitter is properly embedded
        If No: Re-install the transmitter
        If Yes: Confirm with the user if he/she has taken water activities?
            If No: The electrode needle is . Exchange sensor
            If Yes: Wait 3 hours and check if the operation of the system resumes to normal. If No: It is the Short circuit. Exchange sensor
    If Current is below 0,1 nA: Check if transmitter is properly embedded
        If No: Re-install the transmitter
        If Yes: The electrode has been disconnected, or no electrode is inserted into the slot. Exchange sensor